AimTo evaluate the effect of different torsional preloads on cyclic fatigue resistance of endodontic rotary instruments constructed from conventional nickel-titanium (NiTi), M-Wire or CM-Wire.MethodologyEighty new size 25, 0.06 taper Mtwo instruments (Sweden & Martina), size 25, 0.06 taper HyFlex CM (Coltene/Whaledent, Inc) and X2 ProTaper Next (Dentsply Maillefer) were used. The Torque and distortion angles at failure of new instruments (n = 10) were measured, and 0% (n = 10), 25%, 50% and 75% (n = 20) of the mean ultimate torsional strength as preloading condition were applied according to ISO 3630-1 for each brand. The twenty files tested for every extent of preload were subjected to 20 or 40 torsional cycles (n = 10). After torsional preloading, the number of cycles to failure was evaluated in a simulated canal with 60° angle of curvature and 5 mm of radius of curvature. Data were analysed using two-way analysis of variance. The fracture surface of each fragment was examined with a scanning electron microscope (SEM). Data were analysed by two-way analyses of variance.ResultsPreload repetitions did not influence the cyclic fatigue of the three brands; however, the 25%, 50% and 75% torsional preloading significantly reduced the fatigue resistance of all instruments tested (P < 0.01, P < 0.001 and P < 0.0001, respectively) except for the HyFlex CM preloaded with 25% of the maximum torsional strength (P > 0.05).ConclusionsTorsional preloads reduced the cyclic fatigue resistance of conventional and treated (M-wire and CM-wire) NiTi rotary instruments except for size 25, 0.06 taper HyFlex CM instruments with a 25% of torsional preloading.

Aim To evaluate the effect of different torsional preloads on cyclic fatigue resistance of endodontic rotary instruments constructed from conventional nickel titanium (NiTi), M-Wire or CM-Wire.Methodology Eighty new size 25, 0.06 taper Mtwo instruments (Sweden & Martina), size 25, 0.06 taper HyFlex CM (Coltene/Whaledent, Inc) and X2ProTaper Next (Dentsply Maillefer) were used. TheTorque and distortion angles at failure of new instruments (n = 10) were measured, and 0% (n = 10), 25%, 50% and 75% (n = 20) of the mean ultimate torsional strength as preloading condition wereapplied according to ISO 3630-1 for each brand. The twenty files tested for every extent of preload were subjected to 20 or 40 torsional cycles (n = 10). After torsional preloading, the number of cycles to failure was evaluated in a simulated canal with 60° angle ofcurvature and 5 mm of radius of curvature. Data were analysed using two-way analysis of variance.The fracture surface of each fragment was examined with a scanning electron microscope (SEM). Data were analysed by two-way analyses of variance. Results Preload repetitions did not influence the cyclic fatigue of the three brands; however, the 25%, 50% and 75% torsional preloading significantly reduced the fatigue resistance of all instruments tested (P < 0.01, P < 0.001 and P < 0.0001, respectively)except for the HyFlex CM preloaded with 25% of the maximum torsional strength (P > 0.05).Conclusions Torsional preloads reduced the cyclic fatigue resistance of conventional and treated (M-wireand CM-wire) NiTi rotary instruments except for size 25, 0.06 taper HyFlex CM instruments with a 25% of torsional preloading.

Influence of cyclic torsional preloading on cyclic fatigue resistance of nickel-titanium instruments

PEDULLA', EUGENIO;LO SAVIO, Fabio Raffaele Emilio An;RAPISARDA, Ernesto;LA ROSA, Guido
2015-01-01

Abstract

AimTo evaluate the effect of different torsional preloads on cyclic fatigue resistance of endodontic rotary instruments constructed from conventional nickel-titanium (NiTi), M-Wire or CM-Wire.MethodologyEighty new size 25, 0.06 taper Mtwo instruments (Sweden & Martina), size 25, 0.06 taper HyFlex CM (Coltene/Whaledent, Inc) and X2 ProTaper Next (Dentsply Maillefer) were used. The Torque and distortion angles at failure of new instruments (n = 10) were measured, and 0% (n = 10), 25%, 50% and 75% (n = 20) of the mean ultimate torsional strength as preloading condition were applied according to ISO 3630-1 for each brand. The twenty files tested for every extent of preload were subjected to 20 or 40 torsional cycles (n = 10). After torsional preloading, the number of cycles to failure was evaluated in a simulated canal with 60° angle of curvature and 5 mm of radius of curvature. Data were analysed using two-way analysis of variance. The fracture surface of each fragment was examined with a scanning electron microscope (SEM). Data were analysed by two-way analyses of variance.ResultsPreload repetitions did not influence the cyclic fatigue of the three brands; however, the 25%, 50% and 75% torsional preloading significantly reduced the fatigue resistance of all instruments tested (P < 0.01, P < 0.001 and P < 0.0001, respectively) except for the HyFlex CM preloaded with 25% of the maximum torsional strength (P > 0.05).ConclusionsTorsional preloads reduced the cyclic fatigue resistance of conventional and treated (M-wire and CM-wire) NiTi rotary instruments except for size 25, 0.06 taper HyFlex CM instruments with a 25% of torsional preloading.
2015
Aim To evaluate the effect of different torsional preloads on cyclic fatigue resistance of endodontic rotary instruments constructed from conventional nickel titanium (NiTi), M-Wire or CM-Wire.Methodology Eighty new size 25, 0.06 taper Mtwo instruments (Sweden &amp; Martina), size 25, 0.06 taper HyFlex CM (Coltene/Whaledent, Inc) and X2ProTaper Next (Dentsply Maillefer) were used. TheTorque and distortion angles at failure of new instruments (n = 10) were measured, and 0% (n = 10), 25%, 50% and 75% (n = 20) of the mean ultimate torsional strength as preloading condition wereapplied according to ISO 3630-1 for each brand. The twenty files tested for every extent of preload were subjected to 20 or 40 torsional cycles (n = 10). After torsional preloading, the number of cycles to failure was evaluated in a simulated canal with 60° angle ofcurvature and 5 mm of radius of curvature. Data were analysed using two-way analysis of variance.The fracture surface of each fragment was examined with a scanning electron microscope (SEM). Data were analysed by two-way analyses of variance. Results Preload repetitions did not influence the cyclic fatigue of the three brands; however, the 25%, 50% and 75% torsional preloading significantly reduced the fatigue resistance of all instruments tested (P &lt; 0.01, P &lt; 0.001 and P &lt; 0.0001, respectively)except for the HyFlex CM preloaded with 25% of the maximum torsional strength (P &gt; 0.05).Conclusions Torsional preloads reduced the cyclic fatigue resistance of conventional and treated (M-wireand CM-wire) NiTi rotary instruments except for size 25, 0.06 taper HyFlex CM instruments with a 25% of torsional preloading.
CM-wire; Cyclic fatigue resistance; NiTi fracture
File in questo prodotto:
File Dimensione Formato  
2014 - Influence of cyclic torsional preloading on cyclic fatigue resistance of nickel – titanium instruments.pdf

solo gestori archivio

Tipologia: Versione Editoriale (PDF)
Dimensione 726.71 kB
Formato Adobe PDF
726.71 kB Adobe PDF   Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.11769/18357
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 54
  • ???jsp.display-item.citation.isi??? 41
social impact