CTDI metric is currently used for standardized measures of dose-output. Recently, however, strengths/weaknesses of CTDI have been debated: (1) 100 mm-long ionization chambers may not be sufficiently long to measure the tails of scattered dose distribution, (2) phantoms used are shorter than an adult torso. Two approaches have been proposed: use longer ionization chambers and phantoms, or using small probes to determine the equilibrium dose ðDeqÞ [1]. The present study proposes a method based on CTDI-in-air metric to calculate Deq following the approach suggested by Li et al. [2]. The method was also validated using radiochromic films.

247. From CTDI-in-air to equilibrium dose

Spampinato, S.;Gueli, A.;Raffaele, L.
2018-01-01

Abstract

CTDI metric is currently used for standardized measures of dose-output. Recently, however, strengths/weaknesses of CTDI have been debated: (1) 100 mm-long ionization chambers may not be sufficiently long to measure the tails of scattered dose distribution, (2) phantoms used are shorter than an adult torso. Two approaches have been proposed: use longer ionization chambers and phantoms, or using small probes to determine the equilibrium dose ðDeqÞ [1]. The present study proposes a method based on CTDI-in-air metric to calculate Deq following the approach suggested by Li et al. [2]. The method was also validated using radiochromic films.
File in questo prodotto:
File Dimensione Formato  
18_PM_CTDI.pdf

solo gestori archivio

Tipologia: Versione Editoriale (PDF)
Dimensione 35.54 kB
Formato Adobe PDF
35.54 kB Adobe PDF   Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.11769/371347
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact