Purpose:To evaluate the reproducibility of the peripapillary retinal nerve fiber layer (RNFL) thickness measurements obtained by Spectralis spectral domain optical coherence tomography (OCT) in normal eyes and eyes with keratoconus. Methods:Peripapillary RNFL thickness was evaluated with Spectralis spectral domain OCT (Heidelberg Engineering, Heidelberg, Germany). Measurements were repeated 3 times during the same visit using the follow-up function. One eye of each participant was randomly selected for statistical analysis. Reproducibility was evaluated using within-subject standard deviation (Sw), coefficient of variation (CV), and intraclass correlation coefficient (ICC). Results:Forty normal eyes and forty eyes with keratoconus were included in the study. The RNFL thickness overall global CV was 1.37% in normal eyes and 2.10% in eyes with keratoconus. The RNFL thickness overall global ICC (95% CI) was 0.987 (0,973-0,994) in normal eyes and 0.964 (0,926-0,984) in eyes with keratoconus. The RNFL thickness overall global Sw ±1.96 standard error was 1,66±0,22 µm in normal eyes and 2,18±0,21 µm in eyes with keratoconus Conclusions:Spectralis OCT shows a very good reproducibility compared with normal eyes for measuring the peripapillary RNFL thickness in keratoconus eyes

Reproducibility Of Retinal Nerve Fiber Layer Thickness Measurements Using Spectral Domain Optical Coherence Tomography In Eyes With Keratoconus

REIBALDI, MICHELE;AVITABILE, Teresio;UVA, Maurizio Giacinto;LONGO, ANTONIO;Andrea Russo
2011-01-01

Abstract

Purpose:To evaluate the reproducibility of the peripapillary retinal nerve fiber layer (RNFL) thickness measurements obtained by Spectralis spectral domain optical coherence tomography (OCT) in normal eyes and eyes with keratoconus. Methods:Peripapillary RNFL thickness was evaluated with Spectralis spectral domain OCT (Heidelberg Engineering, Heidelberg, Germany). Measurements were repeated 3 times during the same visit using the follow-up function. One eye of each participant was randomly selected for statistical analysis. Reproducibility was evaluated using within-subject standard deviation (Sw), coefficient of variation (CV), and intraclass correlation coefficient (ICC). Results:Forty normal eyes and forty eyes with keratoconus were included in the study. The RNFL thickness overall global CV was 1.37% in normal eyes and 2.10% in eyes with keratoconus. The RNFL thickness overall global ICC (95% CI) was 0.987 (0,973-0,994) in normal eyes and 0.964 (0,926-0,984) in eyes with keratoconus. The RNFL thickness overall global Sw ±1.96 standard error was 1,66±0,22 µm in normal eyes and 2,18±0,21 µm in eyes with keratoconus Conclusions:Spectralis OCT shows a very good reproducibility compared with normal eyes for measuring the peripapillary RNFL thickness in keratoconus eyes
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.11769/70255
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact