ELISSA is an array of silicon strip detectors under construction at the ELI-NP facility for measurements of photodissociation reactions using high-brilliance, quasi monoenergetic gamma beams. The detection system consists of 35 single-sided position-sensitive X3 detectors arranged in a cylindrical configuration and eight QQQ3 detectors as end-caps. A batch of forty X3 detectors have been tested at ELI-NP. The energy and position resolution, ballistic deficit, leakage currents, and depletion voltage were measured and analyzed. Measurements of the energy resolution were carried out using two read-out electronic chains, one based on multichannel preamplifiers and another based on multiplexers.
Titolo: | Performance studies of X3 silicon detectors for the future ELISSA array at ELI-NP |
Autori interni: | |
Data di pubblicazione: | 2018 |
Rivista: | |
Handle: | http://hdl.handle.net/20.500.11769/369517 |
Appare nelle tipologie: | 1.1 Articolo in rivista |