Introduction: This laboratory study aimed to evaluate the accumulated hard tissue debris (AHTD) and shaping performance following root canal instrumentation with a novel stainless steel rotary system (Gentlefile; MedicNRG, Kibbutz Afikim, Israel) compared with two single-file nickel-titanium instruments of different kinematics through micro-computed tomography (micro-CT) analysis.Methods: Thirty isthmus-containing mesial roots of human mandibular molars with a curvature of 15 degrees-35 degrees were scanned with micro-CT at an isotropic resolution of 12 mu m and randomly assigned to three groups (n = 10) according to the instruments used for canal preparation: Gentlefile (GF; #23/.04), One Curve (OC; #25/.06; Micro-Mega, Besancon, France), and Reciproc Blue R25 (RB; #25/.08; VDW, Munich, Germany). The AHTD, unprepared canal surface area, volume changes, surface area changes, and transportation were evaluated by comparing preoperative and postoperative micro-CT images. In addition, the time required for canal preparation was recorded. Data were statistically analyzed using the one-way analysis of variance (ANOVA) test or the Kruskal-Wallis test at a significance level of 5%.Results: No significant differences were observed in the percentage of AHTD or in the unprepared canal surface area between the three groups (P > .05). The RB group demonstrated a significantly greater percentage increase in volume and surface area than the GF group (P < .05). The GF and OC groups showed significantly less transportation at 3 mm from the apex than the RB group (P < .001). The OC group required significantly less time for instrumentation than the RB and GF groups (P < .05).Conclusions: Instrumentation with GF (#23/.04), OC, and RB yielded similar levels of AHTD and unprepared surface area in isthmus-containing curved canals. GF achieved this with less dentin removal and apical transportation, at the expense of a longer preparation time.

Accumulated Hard Tissue Debris and Root Canal Shaping Profiles Following Instrumentation with Gentlefile, One Curve, and Reciproc Blue

Pedullà Eugenio
2023-01-01

Abstract

Introduction: This laboratory study aimed to evaluate the accumulated hard tissue debris (AHTD) and shaping performance following root canal instrumentation with a novel stainless steel rotary system (Gentlefile; MedicNRG, Kibbutz Afikim, Israel) compared with two single-file nickel-titanium instruments of different kinematics through micro-computed tomography (micro-CT) analysis.Methods: Thirty isthmus-containing mesial roots of human mandibular molars with a curvature of 15 degrees-35 degrees were scanned with micro-CT at an isotropic resolution of 12 mu m and randomly assigned to three groups (n = 10) according to the instruments used for canal preparation: Gentlefile (GF; #23/.04), One Curve (OC; #25/.06; Micro-Mega, Besancon, France), and Reciproc Blue R25 (RB; #25/.08; VDW, Munich, Germany). The AHTD, unprepared canal surface area, volume changes, surface area changes, and transportation were evaluated by comparing preoperative and postoperative micro-CT images. In addition, the time required for canal preparation was recorded. Data were statistically analyzed using the one-way analysis of variance (ANOVA) test or the Kruskal-Wallis test at a significance level of 5%.Results: No significant differences were observed in the percentage of AHTD or in the unprepared canal surface area between the three groups (P > .05). The RB group demonstrated a significantly greater percentage increase in volume and surface area than the GF group (P < .05). The GF and OC groups showed significantly less transportation at 3 mm from the apex than the RB group (P < .001). The OC group required significantly less time for instrumentation than the RB and GF groups (P < .05).Conclusions: Instrumentation with GF (#23/.04), OC, and RB yielded similar levels of AHTD and unprepared surface area in isthmus-containing curved canals. GF achieved this with less dentin removal and apical transportation, at the expense of a longer preparation time.
2023
Microcomputed tomography
reciprocation
rotation
transportation
File in questo prodotto:
File Dimensione Formato  
Accumulated Hard Tissue Debris.pdf

solo gestori archivio

Tipologia: Versione Editoriale (PDF)
Licenza: NON PUBBLICO - Accesso privato/ristretto
Dimensione 3.96 MB
Formato Adobe PDF
3.96 MB Adobe PDF   Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.11769/625869
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 4
  • ???jsp.display-item.citation.isi??? 3
social impact