Background: Ocular emergencies account for 1.5-3% of emergency department (ED) visits and require urgent diagnosis to prevent serious complications. Ultrasonography is a crucial, non-invasive diagnostic tool for these conditions but traditionally lacks portability and integration with modern electronic smart devices. The purpose of this study was to assess the accuracy and performance of a new handheld ultrasound device in comparison to a conventional cart-based sonographic machine in patients attending to the ED for vitreo-retinal diseases. (2) Methods: three specialists in ophthalmology, with at least 4-year experience in vitreo-retinal diseases and eye ultrasound, evaluated images of 50 eyes with both portable and traditional ultrasound probes. Each specialist made the diagnosis based on the images captured with both probes and then rated their overall image quality and confidence of diagnosis with a five-point Likert scale. The concordance of diagnosis between the two probes was evaluated. (3) Results: The sample comprised 42 patients. Twenty (40%) healthy eyes and thirty eyes with the following vitreo-retinal interface conditions were examined: 12 retinal detachment (24%), 8 vitreous hemorrhage (16%), and 10 posterior vitreous detachment (20%). The overall accuracy of the two devices appeared to be comparable (70.7% vs. 69.3%). The Butterfly iQ+ probe showed similar sensitivity in retinal detachment diagnosis (91.7% vs. 94.4% of the Accutome B-scan Pro), while it showed poor performance in diagnosing posterior vitreous detachment (sensitivity = 27.2%); (4) Conclusions: The Butterfly iQ+ device demonstrated high sensitivity in the diagnosis of retinal detachment. Significant adjustments are still needed to improve the resolution of the vitreous body.

Comparing New Smartphone-Connected Handheld Ultrasound Device vs. Traditional Ultrasound in Vitreo-Retinal Disease Diagnosis

Rubegni, Giovanni;Russo, Andrea;Castellino, Niccolò;Gagliano, Caterina;Avitabile, Teresio
2024-01-01

Abstract

Background: Ocular emergencies account for 1.5-3% of emergency department (ED) visits and require urgent diagnosis to prevent serious complications. Ultrasonography is a crucial, non-invasive diagnostic tool for these conditions but traditionally lacks portability and integration with modern electronic smart devices. The purpose of this study was to assess the accuracy and performance of a new handheld ultrasound device in comparison to a conventional cart-based sonographic machine in patients attending to the ED for vitreo-retinal diseases. (2) Methods: three specialists in ophthalmology, with at least 4-year experience in vitreo-retinal diseases and eye ultrasound, evaluated images of 50 eyes with both portable and traditional ultrasound probes. Each specialist made the diagnosis based on the images captured with both probes and then rated their overall image quality and confidence of diagnosis with a five-point Likert scale. The concordance of diagnosis between the two probes was evaluated. (3) Results: The sample comprised 42 patients. Twenty (40%) healthy eyes and thirty eyes with the following vitreo-retinal interface conditions were examined: 12 retinal detachment (24%), 8 vitreous hemorrhage (16%), and 10 posterior vitreous detachment (20%). The overall accuracy of the two devices appeared to be comparable (70.7% vs. 69.3%). The Butterfly iQ+ probe showed similar sensitivity in retinal detachment diagnosis (91.7% vs. 94.4% of the Accutome B-scan Pro), while it showed poor performance in diagnosing posterior vitreous detachment (sensitivity = 27.2%); (4) Conclusions: The Butterfly iQ+ device demonstrated high sensitivity in the diagnosis of retinal detachment. Significant adjustments are still needed to improve the resolution of the vitreous body.
2024
POCUS
telemedicine
vitreo-retinal diseases
File in questo prodotto:
File Dimensione Formato  
diagnostics-14-01961.pdf

accesso aperto

Tipologia: Versione Editoriale (PDF)
Licenza: Creative commons
Dimensione 1.74 MB
Formato Adobe PDF
1.74 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.11769/637117
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 0
  • ???jsp.display-item.citation.isi??? 0
social impact