Purpose: The Mobius3D system was validated as a modern secondary check dosimetry system. In particular, our objective has been to assess the suitability of the M3D as pre-treatment patient-specific Quality Assurance (QA) tool for Stereotactic Radiosurgery (SRS) HyperArc (HA) treatments. We aimed to determine whether Mobius3D could safely replace the measurements-based patient-specific QA for this type of treatment. Methods: 30 SRS HA treatment plans for brain were selected. The dose distributions, calculated by Mobius and our routinely used algorithm (AcurosXB v.15.6), were compared using gamma analysis index and DVH parameters based on the patient's CT dataset. All 30 plans were then delivered across the ionization chamber in a homogeneous phantom and the measured dose was compared with both M3D and TPS calculated one. The plans were delivered and verified in terms of PSQA using the electronic portal imaging device (EPID) with Portal Dosimetry (PD) and myQA SRS (IBA Dosimetry) detector. Plans that achieved a global gamma passing rate (GPR) ≥ 97% based on 2%/2 mm criteria, with both Mobius3D and the conventional methods were evaluated acceptable. Finally, we assessed the capability of the M3D system to detect errors related to the position of the Multi-Leaf Collimator (MLC) in comparison to the analyzed measurement-based systems. Results: No relevant differences were observed in the comparison between the dose calculated on the CT-dataset by M3D and the TPS. Observed discrepancies are imputable to different used algorithms, but no discrepancies related to goodness of plans have been found. Average differences between calculated (M3D and TPS) vs measured dose with ionization chamber were 2.5% (from 0.41% to 3.2%) and 1.81% (from 0.66% to 2.65%), for M3D and TPS, respectively. All plans passed with a gamma passing rate > 97% using conventional PSQA methods with a gamma criterion of 2% dose difference and 2 mm distance-to-agreement. The average gamma passing rate for the M3D system was determined to be 99.4% (from 97.3% to 100%). Results from this study also demonstrated Mobius has better error detectability than conventional measurement-based systems. Conclusion: Our study shows Mobius3D could be a suitable alternative to conventional measured based QA methods for SRS HyperArc treatments.

Is it still necessary to perform measured based pre-treatment patient-specific QA for SRS HyperArc treatments?

D'Anna A.;Stella G.;
2023-01-01

Abstract

Purpose: The Mobius3D system was validated as a modern secondary check dosimetry system. In particular, our objective has been to assess the suitability of the M3D as pre-treatment patient-specific Quality Assurance (QA) tool for Stereotactic Radiosurgery (SRS) HyperArc (HA) treatments. We aimed to determine whether Mobius3D could safely replace the measurements-based patient-specific QA for this type of treatment. Methods: 30 SRS HA treatment plans for brain were selected. The dose distributions, calculated by Mobius and our routinely used algorithm (AcurosXB v.15.6), were compared using gamma analysis index and DVH parameters based on the patient's CT dataset. All 30 plans were then delivered across the ionization chamber in a homogeneous phantom and the measured dose was compared with both M3D and TPS calculated one. The plans were delivered and verified in terms of PSQA using the electronic portal imaging device (EPID) with Portal Dosimetry (PD) and myQA SRS (IBA Dosimetry) detector. Plans that achieved a global gamma passing rate (GPR) ≥ 97% based on 2%/2 mm criteria, with both Mobius3D and the conventional methods were evaluated acceptable. Finally, we assessed the capability of the M3D system to detect errors related to the position of the Multi-Leaf Collimator (MLC) in comparison to the analyzed measurement-based systems. Results: No relevant differences were observed in the comparison between the dose calculated on the CT-dataset by M3D and the TPS. Observed discrepancies are imputable to different used algorithms, but no discrepancies related to goodness of plans have been found. Average differences between calculated (M3D and TPS) vs measured dose with ionization chamber were 2.5% (from 0.41% to 3.2%) and 1.81% (from 0.66% to 2.65%), for M3D and TPS, respectively. All plans passed with a gamma passing rate > 97% using conventional PSQA methods with a gamma criterion of 2% dose difference and 2 mm distance-to-agreement. The average gamma passing rate for the M3D system was determined to be 99.4% (from 97.3% to 100%). Results from this study also demonstrated Mobius has better error detectability than conventional measurement-based systems. Conclusion: Our study shows Mobius3D could be a suitable alternative to conventional measured based QA methods for SRS HyperArc treatments.
2023
HyperArc
Mobius3D
PSQA
SRS
File in questo prodotto:
File Dimensione Formato  
2023_Cavalli et al_J Applied Clin Med Phys_QA .pdf

accesso aperto

Tipologia: Versione Editoriale (PDF)
Licenza: Creative commons
Dimensione 1.54 MB
Formato Adobe PDF
1.54 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.11769/576869
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 1
  • ???jsp.display-item.citation.isi??? 1
social impact