Aim: Amplitude-integrated electroencephalography (aEEG)'s accuracy compared to conventional electroencephalography (cEEG) has not been fully established. The aim of our study was to conduct a systematic review on the sensitivity of the aEEG for neonatal seizure detection. Methods: Studies from PubMed and Google Scholar databases comparing recordings of cEEG and aEEG in newborns were included according to the PRISMA method. A quality assessment using the QUADAS-2 tool was provided. A random-effect model was used to account for different sources of variations among studies. Publication biases were represented by a funnel plot, and funnel plot symmetry was assessed. Results: Fourteen studies were reported; sensitivity of each diagnostic tool used (single-channel aEEG, two-channel aEEG, two-channel aEEG plus raw trace EEG) was compared to that of the gold-standard cEEG and to those of the other methods used. Overall sensitivity of the aEEG ranged from 31.25% to 90%. Conclusion: Our study provides evidence that sensitivity of aEEG varies significantly and that seizure detection rate is lower than that of cEEG. The two-channel aEEG with raw trace EEG shows a high sensitivity and might represent a valid alternative to the cEEG in the setting of neonatal intensive care units (NICUs).
aEEG vs cEEG's sensivity for seizure detection in the setting of neonatal intensive care units: A systematic review and meta-analysis
Falsaperla, Raffaele;Scalia, Bruna;Giaccone, Fabiola;Pulvirenti, Alfredo;Ruggieri, Martino
2022-01-01
Abstract
Aim: Amplitude-integrated electroencephalography (aEEG)'s accuracy compared to conventional electroencephalography (cEEG) has not been fully established. The aim of our study was to conduct a systematic review on the sensitivity of the aEEG for neonatal seizure detection. Methods: Studies from PubMed and Google Scholar databases comparing recordings of cEEG and aEEG in newborns were included according to the PRISMA method. A quality assessment using the QUADAS-2 tool was provided. A random-effect model was used to account for different sources of variations among studies. Publication biases were represented by a funnel plot, and funnel plot symmetry was assessed. Results: Fourteen studies were reported; sensitivity of each diagnostic tool used (single-channel aEEG, two-channel aEEG, two-channel aEEG plus raw trace EEG) was compared to that of the gold-standard cEEG and to those of the other methods used. Overall sensitivity of the aEEG ranged from 31.25% to 90%. Conclusion: Our study provides evidence that sensitivity of aEEG varies significantly and that seizure detection rate is lower than that of cEEG. The two-channel aEEG with raw trace EEG shows a high sensitivity and might represent a valid alternative to the cEEG in the setting of neonatal intensive care units (NICUs).File | Dimensione | Formato | |
---|---|---|---|
Acta Paediatrica - 2022 - Falsaperla - aEEG vs cEEG s sensivity.pdf
solo gestori archivio
Tipologia:
Versione Editoriale (PDF)
Licenza:
NON PUBBLICO - Accesso privato/ristretto
Dimensione
1.36 MB
Formato
Adobe PDF
|
1.36 MB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.